10 Beurer GL50 evo mg/dLDisposal Warning• It is essential to comply with the generally applicable safety precautions for handling blood when dis-pos
Beurer GL50 evo mg/dL 113 description of units and accessories3.1 Blood glucose monitorAn overview of the monitor1 2 31 Lancing device 2 Plug-in U
12 Beurer GL50 evo mg/dL3.2 Lancing device and lancet needles1 AST cap (transparent) 2 Cap3 Protective lancet disc4 Sterile lancet needle5 Lan
Beurer GL50 evo mg/dL 133.5 Test strips Front Rear 1 Gap for blood input2 Handle3 ContactsYou can identify the rear by the contact tracks.Insert t
14 Beurer GL50 evo mg/dL• The test strips expire three months after the box is opened. Note down the expiry date (date of open-ing plus three months
Beurer GL50 evo mg/dL 154.2 Making and changing basic settings1Setting the date and timeTo activate the device, set the sliding switch for on/o to s
16 Beurer GL50 evo mg/dL3Setting the timerYou can set a timer on the blood glucose monitor. An acoustic signal sounds when the timer elapses. To set
Beurer GL50 evo mg/dL 176Switching the ketone warning on/oOn the blood glucose monitor, you can switch a ketone warning on and o.1. Switch on the b
18 Beurer GL50 evo mg/dL5 taking measurements Warning• If the protective disc on a lancet needle has already been removed, do not use the lancet nee
Beurer GL50 evo mg/dL 195.2 Preparing the lancing device for taking a sample1Remove capSecurely hold the monitor with one hand on the lancing device
20 Beurer GL50 evo mg/dL5Select the penetration depthYou can set seven di erent penetration depths on the lancing device using the dial with raised
Beurer GL50 evo mg/dL 213The device switches on automatically After you have inserted the test strip, the blood glucose monitor switches on automatic
22 Beurer GL50 evo mg/dL NoteOnly take blood from other areas of the body at the following times:• On an empty stomach (more than two hours after mo
Beurer GL50 evo mg/dL 235.4 Reading the result and labelling measurementsRead the resultAs soon as the blood input gap is filled with sucient blood,
24 Beurer GL50 evo mg/dL5.5. Post-processing and disposal1Remove test stripRemove the test strip from the device and carefully dispose of it in accor
Beurer GL50 evo mg/dL 25 Warning• If you suspect that the blood glucose results are incorrect, first repeat the test and, if applicable, perform a fu
26 Beurer GL50 evo mg/dL5.7 Functional check using control solutionThe control solution is used to test the entire blood glucose monitoring system. T
Beurer GL50 evo mg/dL 273Drip control solution on surfaceChoose a clean surface to carry out the func-tional test correctly. Shake the control soluti
28 Beurer GL50 evo mg/dLIf measured values are outside the specified range, check the following possible causes:Cause Action• The first drop of contro
Beurer GL50 evo mg/dL 296 measurement memoryFor each measurement, your blood glucose value is automatically saved with the date and time unless “CONT
Beurer GL50 evo mg/dL 31 Getting to know your instrument ...4Getting
30 Beurer GL50 evo mg/dL6.2 Displaying average blood glucose valuesYou can display the average measured blood glucose value from the last 7, 14, 30 a
Beurer GL50 evo mg/dL 31Evaluating measured values on the PC1The blood glucose monitor must be switched o. Insert the monitor’s USB connector into a
32 Beurer GL50 evo mg/dL NoteThe blood glucose monitor is made of precision components. The accuracy of the measurements and service life of the devi
Beurer GL50 evo mg/dL 33Problem: device does not switch onCause SolutionBattery empty. Charge the blood glucose monitor via the USB port on a compute
34 Beurer GL50 evo mg/dL9 technical specificationsDimensions (L x W x H) 123 x 29 x 16 mmWeight 36 gPower supply Lithium-ion rechargeable battery, 16
Beurer GL50 evo mg/dL 35Only a small amount of blood is required (0.6 microlitres) and measurement takes approx. five sec-onds. The test strip detects
36 Beurer GL50 evo mg/dLResults of the repeat precision measurementsSample Venous blood (mg/dL)Grand mean value (mg/dL)Pooled standard de-viationPoo
Beurer GL50 evo mg/dL 37Results of the system accuracy for combined glucose concentrations between 34.4 mg/dL (1.9 mmol/L) and 442.8 mg/dL (24.6 mmol
Concentration of tested substances50-100 mg/dL(2.8-5.6 mmol/L)250-350 mg/dL(13.9-19.4 mmol/L)Acetaminophen 7 mg/dL (0.46 mmol/L) 8.1 5.3%Ascorbic aci
Beurer GL50 evo mg/dL 3912 guarantee and customer serviceWarrantyWe oer a 3-year product warranty against material and production faults. The warran
4 Beurer GL50 evo mg/dL1 getting to knoW your instrumentDear customer,Thank you for choosing one of our products. Our name stands for high-quality, t
GL50_mg/dL_1014_GB Subject to errors and changes
Beurer GL50 evo mg/dL 51.1 Delivery scope, replacements and accessoriesCheck that the set packaging has not been tampered with and make sure that all
6 Beurer GL50 evo mg/dLFollow-up purchasesYou can also obtain test strips, control solution and lancet needles without a prescription.Item REF50 test
Beurer GL50 evo mg/dL 71.3 Signs and symbolsThe symbols on the packaging, type plate of the blood glucose monitor and accessories represent the follo
8 Beurer GL50 evo mg/dL2 Warnings and safety notesRisk of infectionAll components of the blood glucose monitor and its accessories may come into cont
Beurer GL50 evo mg/dL 9• An excessively high or low hematocrit value (proportion of red blood cells) may lead to incorrect measurements. In the event